Description
Temperature:
•Working-temperature range 5 °C above ambient temperature up to +50 °C, Standard sterilization program: 60 minutes at 180°C (without removing the sensors)
•Setting temperature range +18 to +50 °C
•Setting resolution temperature 0.1 °C
•Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
•Temperature variation in chamber at 37 °C (to DIN 12880:2007-05) +/- 0.3 °C
•Temperature fluctuations with time at 37 °C (to DIN 12880:2007-05) +/- 0.1 °C
Humidity:
•Humidity control (standard) Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
•Setting range active
•humidity control (with option K7) 40 to 97 % rh and rh-Off
•Setting resolution humidity 0.5 % rh
Control of Standard Components:
•CO2 control Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
•Adjustment range CO2 0 to 20 % CO2
•Setting accuracy CO2 0.1%
•Variation in time CO2 +/- 0.2 % CO2
•Adjustment range O2 (with option T6) 1 to 20 % O2
•Setting accuracy O2 0.1 % O2
Control Technology:
•ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
•Language setting German, English, Spanish, French, Polish, Czech, Hungarian
•Function SetpointWAIT the process time does not start until the set temperature is reached
•Adjustable parameters temperature (Celsius or Fahrenheit), CO2, timer
Communication:
•Interface Ethernet LAN, USB
•Documentation program stored in case of power failure
•Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
Safety:
•AutoSAFETY additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
•Autodiagnostic system integral fault diagnostics for temperature and CO2
•Alarm visual and acoustic