Memmert Constant Climate Chamber 9 cu ft

Quote

Memmert Constant Climate Chamber (HPPeco) – this programmable model is the 230 volt or 115 volt, Peltier-technology seamlessly heats and cools, with TwinDISPLAY digital controller. HPPeco humidity range is 10 to 85% rh without light (upto 90% with light). Each individual appliance is tested according to the requirements of DIN 12 880: 2007-05. Includes (2) Stainless Steel Grid Shelves, Atmocontrol Software, Atmocontrol Manual, Operation manual, Calibration Certificate and Power Cord – energy efficient, precise.

Description

Temperature:

โ€ข Setting temperature range with humidity: +5 to +70 ยฐC
โ€ข Setting temperature range without humidity: 0 to +70 ยฐC
โ€ข Working temperature range with humidity: +5 (at least 20 below ambient temperature) to +70 ยฐC
โ€ข Working temperature range without humidity: 0 (at least 20 below ambient temperature) to +70 ยฐC
โ€ข Setting accuracy temperature 0.1 ยฐC
โ€ข Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity:

โ€ข Setting range humidity 10 – 90 % rh
โ€ข Humidity supply with distilled water from external tank by self-priming pump
โ€ข Humidification by hot steam generator
โ€ข Dehumidification by cold trap using the Peltier technology
โ€ข Setting accuracy humidity 0.5 % rh

Control Technology:

โ€ข ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
โ€ข Language setting German, English, Spanish, French, Polish, Czech, Hungarian
โ€ข Adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, timer
โ€ข Timer digital backwards counter with target time setting, adjustable from 1 minute to 99 days
โ€ข Function SetpointWAIT the process time does not start until the set temperature is reached
โ€ข Calibration three freely selectable values each, temperature and humidity

Communication:

โ€ข Communication interface Ethernet LAN, USB
โ€ข Documentation program stored in case of power failure
โ€ข Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port

Safety:

โ€ข Temperature control over- and under temperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
โ€ข AutoSAFETY additionally integrated over- and under temperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or under temperature, heating function is switched off in case of overtemperature, cooling function in case of under temperature
โ€ข Autodiagnostic system integral fault diagnostics for temperature and humidity control
โ€ข Alarm visual and acoustic

Heating Concept:

โ€ข Energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)

Stainless steel interior:
โ€ข Dimensions w(A) x h(B) x d(C): 25.2 x 31.5 x 19.7 inches (d less 1.3โ€ for fan - Peltier)
โ€ข Volume 256 l / 9 cu. Ft.
โ€ข Max. number of shelves 9
โ€ข Max. loading of chamber 440 lbs
โ€ข Max. loading per shelf 44lbs Textured stainless steel casing:
โ€ข Dimensions w(D) x h(E) x d(F): 32.4 x 46.6 x 25.8 inches (d +2.2โ€ door handle))
โ€ข Housing rear zinc-plated steel Electrical data:
โ€ข Voltage Electrical Load 230 V (ยฑ 10%), 50/60 Hz approx. 700 W / 3.1 A
โ€ข Voltage Electrical Load 115 V (ยฑ 10%), 50/60 Hz approx. 700 W / 6.1 A
No information available, please contact us for information