Description
Temperature:
•Setting temperature range +20 to +80 °C
•Working temperature range min. 10°C above ambient up to +80°C
•Setting resolution temperature 0.1 °C
•Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Control Technology:
•ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
•Lanuage setting German, English, Spanish, French, Polish, Czech, Hungarian
•Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
•Function SetpointWAIT the process time does not start until the set temperature is reached
•Calibration three freely selectable temperature values adjustable parameters temperature (Celsius or Fahrenheit), fan speed, air flap position, timer
•Sterilization fixed sterilization program (4 hours/160°C) for sterilization of working chamber, not for sterilizing the load
Ventilation:
•Fan forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
•Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
•Vent connection with restrictor flap
Communication:
•Documentation program stored in case of power failure
Safety
•Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
•Temperature control overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
•AutoSAFETY additionally integrated over- and undertemperature monitor “ASF”, automatically following the setpoint
•value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
•Autodiagnostic system for fault analysis
•Alarm visual and acoustic