Description
Temperature:
•Setting temperature range +20 to +80 °C
•Working temperature range min. 10°C above ambient up to +80°C
•Setting resolutiontemperature0.1 °C
•Temperature sensor 1 Pt100 sensor DIN class A in 4-wire-circuit
Control Technology:
•Language setting German, English, Spanish, French, Polish, Czech, Hungarian
•ControlCOCKPIT SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display
•Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
•Function SetpointWAIT the process time does not start until the set temperature is reached
•Calibration three freely selectable temperature values
•adjustable parameters temperature (Celsius or Fahrenheit), fan speed, air flap position, timer
Ventilation:
•Fan forced air circulation by quiet air turbine, adjustable in 10 % steps
•Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
•Vent connection with restrictor flap
Communication:
•Documentation settings stored in case of power failure
•Data logging AtmoCONTROL software for reading out, managing and organizing the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).
Safety
•Temperature control adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
•Autodiagnostic system for fault analysis